Recall archive
31 recalls · page 2 of 2
Reported Mar 13, 2019
AXIUS Blower Mister Product Code/REF Number: CB-1000 The Axius Blower Mister is intended to clear an anastomotic site…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Maquet Cardiovascular, LLC
Matched: Brand — Potential lack of carbon dioxide (CO2) flow that may result in a procedural delay during off-pump coronary artery bypass (OPCAB).
FDA (openFDA device enforcement)
Reported Mar 22, 2017
VASOVIEW HEMOPRO VH-3500 ENDOSCOPIC VESSEL HARVESTING SYSTEM Product Usage: The VASOVIEW HEMOPRO VH-3500 is designed…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Maquet Cardiovascular, LLC
Matched: Brand · Reason for recall — Maquet has received several complaints involving the VASOVIEW HEMOPRO VH-3500 which involve the device exhibiting the failure mode of intermittent or…
FDA (openFDA device enforcement)
Reported Nov 5, 2014
Acrobat V Vacuum Off-Pump System OM-9100S Acrobat SUV Vacuum Off-Pump System OM-9000S Axius Xpose 3 Access Device…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Maquet Cardiovascular, LLC
Matched: Brand — An internal finding identified the presence of pinholes/cuts in some of the Tyvek pouches of the vacuum tubing sets included as an accessory to…
FDA (openFDA device enforcement)
Reported Dec 18, 2013
Vasoview Hemopro Endoscopic Vessel Harvesting System Maquet Cardiovascular LLC 45 Barbour Pond Drive Wayne, NJ 07470…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Maquet Cardiovascular, LLC
Matched: Brand · Product · Product details · Reason for recall — Maquet has received an increase in the number of complaints for "self-activated/remains activated/overheats" for the Hemopro VH-30000, "Hemopro 1"
FDA (openFDA device enforcement)
Reported Oct 9, 2013
Ultima OPCAB System, Sterile, Rx Only, Product Usage: Intended for use during performance of minimally invasive cardiac…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Maquet Cardiovascular, LLC
Matched: Brand · Reason for recall — MAQUET has identified a potential issue with the ULTIMA OPCAB (part numbers OM-2003S & OM-2001D). Labeling for the 2 blade lots had been incorrectly…
FDA (openFDA device enforcement)
Reported Apr 24, 2013
MAQUET CARDIOHELP Support System: CARDIOHELP-I Software (up to version 3.3.0.0) Product Usage: The Cardiohelp System is…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Maquet Cardiovascular, LLC
Matched: Brand · Product · Product details · Reason for recall — It has come to the attention of MAQUET that there have been episodes of brief unexpected shutdown, followed by an automatic device restart of the…
FDA (openFDA device enforcement)
Reported Jan 16, 2013
CARDIOSAVE Intra-Aortic Balloon Pump An electromechanical system used to inflate and deflate intra-aortic balloons. It…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Maquet Cardiovascular, LLC
Matched: Brand — In the Cardiosave Intra-aortic Balloon Pump, there is a possibility that the coiled cord assembly, a prominent connector cord between the Cardiosave…
FDA (openFDA device enforcement)