Recall archive
59 recalls · page 3 of 3
Reported Oct 8, 2014
The Symbia T Series consists of the T16, T6, T2, and T. The Symbia T integrates state-of-the-art SPECT and high quality…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Medical Solutions USA, Inc.
Matched: Brand · Reason for recall — Siemens Medical Solutions USA, Inc. is performing a field correction because we recently detected a rubber isolator with unfavorable aging…
1 more product in this recall
FDA (openFDA device enforcement)
Reported Sep 10, 2014
V5Ms Transesophageal transducers; V5Ms trans-esophageal echocardiograph (TEE) ultrasound transducer is intended…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Medical Solutions USA, Inc.
Matched: Brand — Reports of deterioration of material covering the articulating section of the V5Ms transesophageal transducer.
FDA (openFDA device enforcement)
Reported Jul 23, 2014
e.cam camera system. Emission Computed Tomography System used to detect or image the distribution of radionuclides in…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Medical Solutions USA, Inc.
Matched: Brand · Reason for recall — Siemens Medical Solutions USA, Inc. has received reports indicating the collimator cart rear casters may become loose.
1 more product in this recall
FDA (openFDA device enforcement)
Reported Jul 9, 2014
ACUSON S Family Ultrasound Systems at software versions VC25D, VC30A and VC30B; (S2000, S3000 and S1000 systems)…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Medical Solutions USA, Inc.
Matched: Brand · Product details — There is a potential measurement error with the Auto-Stats (auto statistics) measurement tool on the ACUSON S Family ultrasound system.
FDA (openFDA device enforcement)
Reported Jun 4, 2014
ACUSON SC2000 Ultrasound Systems with software version 3.5 (VA35); ultrasound imaging system.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Medical Solutions USA, Inc.
Matched: Brand — Software issues with ACUSON SC2000 ultrasound with software version 3/5 (VA35) - may incorrectly display the color flow data in the wrong position…
1 more product in this recall
FDA (openFDA device enforcement)
Reported Apr 30, 2014
Siemens ACUSON X700 Ultrasound Systems with software version 1.0.00, 1.0.01, 1.0.02, 1.1.00 and 1.1.02. ultrasound…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Medical Solutions USA, Inc.
Matched: Brand · Product · Product details — When fourSight 4D ultrasound imaging, 3-Scrape real time 3D imaging, or Anatomical M-mode are used during an Obstetric or Gynecology study, data from…
FDA (openFDA device enforcement)
Reported Apr 9, 2014
Siemens Medical Solutions, ACUSON P300 Ultrasound system with P300 PA 230E cardiac probe (Esaote product 7348).…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Medical Solutions USA, Inc.
Matched: Brand · Product · Product details — ACUSON P300 ultrasound systems using the PA230 transducer with a user-defined preset which includes the 100% setting for output power could cause the…
FDA (openFDA device enforcement)
Reported Feb 5, 2014
ACUSON SC2000 Ultrasound System; Power Input: 100-240VAC, 1600W Max, 50/60 Hz. Manufacturer: Siemens Medical Solutions…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Medical Solutions USA, Inc.
Matched: Brand · Product · Product details — The user interface assembly on the ACUSON SC2000 may become loose with the potential for the entire module to fall off of the ultrasound system.
1 more product in this recall
FDA (openFDA device enforcement)
Reported Oct 30, 2013
Acuson SC2000 volume imaging ultrasound systems at software version 3.0. Mfg name: Siemens Medical Solutions USA, Inc.…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Medical Solutions USA, Inc.
Matched: Brand · Product · Product details — Software issue results in inaccurate 3D image may result in ablation in wrong location.
FDA (openFDA device enforcement)
Reported Apr 17, 2013
Siemens brand ACUSON SC2000 Ultrasound System, Model 10433816, hardware that contains version 4 of the MPI board…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Medical Solutions USA, Inc.
Matched: Brand · Product · Product details — When using the V5M transducer, the display of the temperature on the ACUSON SC2000 system can be lower than the actual temperature of the V5M…
FDA (openFDA device enforcement)
Reported Mar 20, 2013
ACUSON S2000 Automated Breast Volume Scanner (ABVS); Model number 10042541 Distributed by Siemens Medical Solutions USA…
Moderate riskThis product could cause temporary or reversible health problems — stop using it.Siemens Medical Solutions USA, Inc.
Matched: Brand · Product · Product details — On some ACUSON S2000 ABVS systems, the inside of the transducer pod may contain sharp edges.
FDA (openFDA device enforcement)
RecallNow is not a government agency. Every notice above is published by the FDA, USDA FSIS, or CPSC and links back to its source. Back to the look-up