Reported May 31, 2024 · Announced

Medline Industries LP recalls Medline SubG Endotracheal Tubes with Subglottic Suction and kits that contain these devices due to The inflation tube and other…

Medline Industries LP

Risk not classifiedThe agency has not ranked the risk yet — treat the notice as a caution.

What you should do

Stop using the product and follow the return or refund instructions in the firm's announcement.

Is this the product you have?

  • SubG Endotracheal Tubes with Subglottic Suction and
  • kits that contain these devices BrandMedline. Medical Devices

Why it was recalled

The inflation tube and other device components have become detached and/or torn from the main tube, resulting in moisture buildup, loss of pressure, or inability to inflate.

The agency has not ranked the risk yet — treat the notice as a caution.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Medline Industries LP recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration

Read the FDA (recalls & safety alerts) notice

Data as of Sep 19, 5:08 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.