Reported Oct 9, 2024 · Ongoing

MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755; MiniMed 700G Insulin Pump, REF:…

Medtronic MiniMed, Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Medtronic MiniMed, Inc. · 18000 Devonshire St · Northridge, CA · 91325-1219

Is this the product you have?

  • MiniMed 630G Insulin Pump
  • REF
    • MMT-1515
    • MMT-1714
    • MMT-1715
    • MMT-1754
    • MMT-1755
    • MiniMed 700G Insulin Pump
  • REF
    • MMT-1800
    • MMT-1801
    • MMT-1805
    • MMT-1850
    • MMT-1851
Lot codes and dates

7 of 7 products

Products covered by this recall
#BarcodesProduct as published
10064316978239600763000253059MMT-1714/00643169782396, 00763000253059
20064316987384100763000190446006431697527266431698738340076300031759100763000316655763000367053763000316655763000316655037630001904397630003166317630001665197630000725207630003166550176300025680776300009019764316965684076300031668600643169656840007630000901970076300019043900763000316631MMT-1715/00643169873841, 00763000190446, 00643169752726, 643169873834, 00763000317591, 00763000316655, 763000367053, 763000316655, 76300031665503, 763000190439, 763000316631, 763000166519, 763000072520, 76300031665501, 763000256807, 763000090197, 643169656840, 763000316686, 00643169656840, 00763000090197, 00763000190439, 00763000316631
36431697527266431698738417630001904467630001015277630003175910076300031759176300036709176300017963276300031762164316987864876300015704376300031759103006431697527267630003175910176300031762103763000317621010076300010152700763000190446MMT-1755/643169752726, 643169873841, 763000190446, 763000101527, 763000317591, 00763000317591, 763000367091, 763000179632, 763000317621, 643169878648, 763000157043, 76300031759103, 00643169752726, 76300031759101, 76300031762103, 76300031762101, 00763000101527, 00763000190446
400763000531157MMT-1801/00763000531157
500763000561680MMT-1805/00763000561680
600763000552541MMT-1850/00763000552541
70076300085473700763000541323MMT-1851/00763000854737, 00763000541323. All serial numbers.
Amount recalled
24,595

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

Why it was recalled

Stop using this product now

Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Medtronic MiniMed, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:57 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.