Reported Oct 11, 2017 · Terminated
Human Alpha-1-Antitrypsin Kit for use on the SPAPlus analyser, Product #NK034.S. The firm name on the label is The…
The Binding Site Group, Ltd.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: The Binding Site Group, Ltd. · 8 Calthorpe Road · Birmingham, N/A · N/A
Is this the product you have?
- Human Alpha-1-Antitrypsin Kit for use on the SPAPlus analyser, Product #NK034.S. The firm name on the label is The Binding Site Group Ltd., Birmingham, UK. Product is designed for the quantitative in-vitro determination of Alpha-1-antitrypsin in human serum using the SPAPLUS turbidimetric analyser. The measurement of Alpha-1-antitrypsin aids in the diagnosis of several conditions including adult cirrhosis of the liver. In addition, Alpha-1-antitrypsin deficiency has been associated with pulmonary emphysema. This test should be used in conjunction with other laboratory and clinical findings.
- Lot codes and dates
Lot #400900Exp. 12/31/2017UDI #05051700016421
- Amount recalled
- 361 kits
Where it was sold
Why it was recalled
The high control for the kit is recovering low and may be out of range for assigned values.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new The Binding Site Group, Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.