Reported Oct 18, 2017 · Terminated
PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
Handicare USA Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Handicare Usa Inc · 10888 Metro Ct · Maryland Heights, MO · 63043-2413
Is this the product you have?
- PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
- Lot codes and dates
A6M0001526A6M0001527A6M0001528A6M0001529A6M0001530A6M0001485A6M0001486A6M0001498
- Amount recalled
- 8 units
Why it was recalled
During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.