Reported Oct 8, 2025 · Ongoing
Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Getinge Disinfection Ab
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Getinge Disinfection Ab · Ljungadalsgatan 11 · Vaxjo, N/A · N/A
Is this the product you have?
- Getinge 88-Series Washer- Disinfector
- Model Number88-5
- Lot codes and dates
- Model Number88-5
- UDI-DI Numbers
0734015370010907340153700116
- Software VersionGD 14080 Rev A. All Lot Numbers with electrical heating
- Amount recalled
- 902 units
Where it was sold
Why it was recalled
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.