Reported Oct 8, 2025 · Ongoing

Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Getinge Disinfection Ab

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Getinge Disinfection Ab · Ljungadalsgatan 11 · Vaxjo, N/A · N/A

Is this the product you have?

  • Getinge 88-Series Washer- Disinfector
  • Model Number88-5
Lot codes and dates
  • Model Number88-5
  • UDI-DI Numbers
    • 07340153700109
    • 07340153700116
  • Software VersionGD 14080 Rev A. All Lot Numbers with electrical heating
Amount recalled
902 units

Where it was sold

Why it was recalled

Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Getinge Disinfection Ab recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:35 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.