Reported Oct 15, 2025 · Ongoing
NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
Noxbox Ltd.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: NOXBOX LTD · Unit 1 · Sittingbourne · N/A
Is this the product you have?
- NOxBOXi Nitric Oxide Delivery System
- Model NumbersNOXBOX-I
- Lot codes and dates
- Model Numbers
(1) NOXBOXI (NBL (Manufacturer))(2) REQNOXBOXI (Linde (U.S. Distributor))
- UDI-DI05060541640009; All serial numbers
- Model Numbers
- Amount recalled
- 1667 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Stop using this product now
The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Noxbox Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.