Reported Oct 12, 2016 · Terminated

RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0) -- Radiation Therapy Treatment Planning System…

Raysearch Laboratories Ab

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: RAYSEARCH LABORATORIES AB · Sveavaegen 9 · Stockholm, N/A · N/A

Is this the product you have?

  • RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0) -- Radiation Therapy Treatment Planning System RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.
Lot codes and dates
  • Build Numbers
    • 2.5.0.144
    • 2.5.1.89
    • 3.0.0.251
    • 3.5.0.16
    • 3.5.1.6
    • 4.0.0.14
    • 4.0.1.4
    • 4.0.2.9
    • 4.0.3.42
    • 4.3.0.14
    • 4.5.0.19
    • 4.5.1.14
    • 4.5.2.7
    • 4.7.0.15
    • 4.7.1.10
    • 4.7.2.5
    • 4.7.4.4
    • 4.7.5.4
    • 5.0.0.37
    • 5.0.1.11 or 5.0.2.35
Amount recalled
USA: 522 units, OUS 846 units

Where it was sold

  • Nationwide

Why it was recalled

RaySearch became aware of the problem as it was discovered in cooperation with a customer experimenting with the dose calculation on a phantom. The correction concerns two issues found with the dose calculation when using a region of interest (ROI) of type Fixation or Support with material override within the patient outline (External ROI) in RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Raysearch Laboratories Ab recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 11, 2019 · Data as of Sep 19, 6:51 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.