Reported Nov 6, 2013 · Terminated

Xoran Minicat00158

Xoran Technologies, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Xoran Technologies, Inc. · 5210 S State Rd · Ann Arbor, MI · 48108-7936

Is this the product you have?

  • Xoran Minicat00158
Lot codes and dates
  • minicat
Amount recalled
N/A

This recall covers 6 products

The agency filed one notice per product. You are reading one of them; the other 5 are below.

Where it was sold

  • Nationwide

Why it was recalled

The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Xoran Technologies, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 11, 2014 · Data as of Sep 19, 5:28 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.