Reported Nov 1, 2017 · Terminated
Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.
Beckman Coulter Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232
Is this the product you have?
- Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.
- Lot codes and dates
- Software version PLC V26R10
- Amount recalled
- 27 units
Why it was recalled
Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.
This product could cause temporary or reversible health problems — stop using it.