Heart Ware(TM) HVAD(TM) System Controller/Controller Kits, Model Numbers: 1403, 1407, and 1420 Product Usage: The…
Heartware, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Heartware, Inc. · 14400 Nw 60th Ave · Miami Lakes, FL · 33014-2807
Is this the product you have?
- Heart Ware(TM) HVAD(TM) System Controller/Controller Kits
- Model Numbers
14031407
- and 1420 Product UsageThe HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned
- Lot codes and dates
- Serial number rangeCON300175 - CON320540
- Amount recalled
- 17,275 units
Why it was recalled
Certain HVAD System Controller units may not meet the labeled standard for protection against water or fluid ingress, specifically they may have the potential to develop hairline cracks at the power ports of the controller housing.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Heartware, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.