DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU Module, DxC…
Beckman Coulter Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232
Is this the product you have?
- DxC 500i Clinical Analyzer Modules
1. DxC 500 AU Module w/ISEDxC 500iPart No. C63522 2. DxC 500 AU ModuleDxC 500iPart No. C63521 3. Access 2 ModuleDxC 500iPart No. C13252
- Lot codes and dates
- Software version 1.3.5 or earlier. 1. UDI-DI 14987666545089 2. UDI-DI 14987666545072 3. UDI-DI 15099590742331
- Amount recalled
- 146 US, 170 OUS
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Beckman Coulter Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.