Reported Oct 26, 2022 · Ongoing

N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN):…

Siemens Healthcare Diagnostics Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: SIEMENS HEALTHCARE DIAGNOSTICS INC · 511 Benedict Ave · Tarrytown, N/A · N/A

Is this the product you have?

  • N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System-
  • IVD Siemens Material Number (SMN)10446299
Lot codes and dates
  • UDI-DI0405686900179VF
  • Lot Number153083; 153085; 153087A; 153090; 153095B; 153099A; 153004B. Future lots may be impacted by the issue but will then be provided with a corresponding instruction note within the reagent package
Amount recalled
10,201 packs

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide

Why it was recalled

N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:37 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.