Reported Nov 5, 2014 · Terminated

The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power…

Hospira Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Hospira Inc. · 275 N Field Dr · Lake Forest, IL · 60045-2579

Is this the product you have?

  • The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.
Lot codes and dates
  • Desktop
  • GemStar Power Supply List Number
    • 13072-05
    • 1)
  • US Distribution Lot Numbers
    • 10115G1
    • 11121G1
    • 11122G1
    • 1123G1
    • 11124G1
    • 12143G1
    • 12144G1
    • 12178G1
    • 15185G1
    • 12195G1
    • 17214G1
    • 18217G1
    • 18219G1
    • 19225G1
    • 20233G1
    • 21242G1
    • 22251G1
    • 23259G1
    • 24271G1
    • 25280G1
    • 26289G1
    • 26290G1
    • 27293G1
    • 27297G1
    • 27300G1
    • 2)
  • Foreign Distribution Lot Numbers
    • 11128G1
    • 12140G1
    • 12142G1
    • 12148G1
    • 14176G1
    • 15187G1
    • 15196G1
    • 16207G1
    • 16208G1
    • 18218G1
    • 19226G1
    • 19229G1
    • 20234G1
    • 21241G1
    • 22255G1
    • 23262G1
    • 24270G1
    • 25276G1
    • 26285G1
    • 27299G1
    • 27301G1
Amount recalled
US Distribution: 4,709 power supplies; Foreign Distribution: 7,425 power supplies

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Stop using this product now

The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Hospira Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 7, 2017 · Data as of Sep 19, 4:47 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.