Reported Oct 26, 2022 · Terminated
cobas 5800 instrument, Material No. 08707464001. Used for automated Polymerase Chain Reaction (PCR) based Nucleic Acid…
Roche Molecular Systems, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Roche Molecular Systems, Inc. · 1080 US Highway 202 S · Branchburg, NJ · 08876-3733
Is this the product you have?
- cobas 5800 instrument, Material No. 08707464001. Used for automated Polymerase Chain Reaction (PCR) based Nucleic Acid Testing (NAT) in laboratory settings.
- Lot codes and dates
- Device Identifier07613336170076 All serial IDs are affected
- Amount recalled
- 179 OUS
Why it was recalled
Reported false positive and invalid results on the affected devices due to anomalous baselines. This may lead to erroneous or delayed diagnoses.