Reported Oct 18, 2023 · Ongoing
Access Thyroglobulin Calibrators
Beckman Coulter, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beckman Coulter, Inc. · 1000 Lake Hazeltine Dr · Chaska, MN · 55318-1037
Is this the product you have?
- Access Thyroglobulin Calibrators
- Lot codes and dates
- UDI/DI 15099590227166
- Serial/Lot Numbers
233806234171234355234443234444234617337940
- Amount recalled
- 5980 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Why it was recalled
It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.