Reported Oct 29, 2014 · Terminated

Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the…

Estill Medical Technologies, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Estill Medical Technologies, Inc · 4144 N Central Expy Ste 260 · Dallas, TX · 75204-3100

Is this the product you have?

  • Ultra Battery 1 ModelTA-UB1
  • NSN6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the Thermal Angel Blood and IV Fluid Infusion Warmer. The TA-UB1 battery was designed to provide sufficient power for the Thermal Angel to warm and deliver one liter of blood or IV fluid infusion to a trauma patient
Lot codes and dates
    • 06-2011
    • 07-2011
    • 08-2011
    • 09-2011
    • 03-2012
    • 04-2012
    • 05-2012
    • 06-2012
    • 07-2012
    • 08-2012
    • 01-2013
    • 03-2013
    • 08-2013
    • 09-2013
    • 10-2013
    • 11-2013
    • 12-2013
    • 01-2014
    • 02-2014
    • 04-2014
    • 07-2012
    • 02-2014
Amount recalled
6,886

Where it was sold

  • Nationwide

Why it was recalled

Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Estill Medical Technologies, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 21, 2015 · Data as of Sep 19, 6:03 PM UTC