Reported Oct 25, 2023 · Ongoing
stryker Tornier Perform Reversed Augmented Glenoid Lateralized Baseplate, 25mm, Offset +3mm, REF DWJ502
Tornier, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Tornier, Inc · 10801 Nesbitt Ave S · Bloomington, MN · 55437-3109
Is this the product you have?
- stryker Tornier Perform Reversed Augmented Glenoid Lateralized Baseplate, 25mm, Offset +3mm, REF DWJ502
- Lot codes and dates
UDI/DI 10846832062017Lot Number 1756123
- Amount recalled
- 30 units
Where it was sold
Why it was recalled
One lot of Tournier Perform Reversed Lateralized Baseplates were found to have been manufactured with 6mm spacers instead of 3mm spacers. This nonconformance prevents the central screw from being able to thread into the baseplate.