Reported Nov 29, 2017 · Terminated
3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)
3M Company - Health Care Business
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: 3M Company - Health Care Business · 3m Center 2510 Conway Ave , B# 275-5-W-6 · Saint Paul, MN · 55144-0001
Is this the product you have?
- 3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)
- Lot codes and dates
- Lot Numbers
17230Exp. 2020 Aug 1817233Exp. 2020 Aug 2117234Exp. 2020 Aug 2217241Exp. 2020 Aug 29
- Lot Numbers
- Amount recalled
- 153 cases (3825 sensors)
Where it was sold
Why it was recalled
During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature.