Reported Oct 16, 2019 · Terminated

Clinical Chemistry, ABX Pentra Lipase CP, ABX Pentra 400/Pentrac400, REF: A11A01631, Model/Part # 1220001631, UDI #…

Horiba Instruments Incorporated

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Horiba Instruments Incorporated · 34 Bunsen · Irvine, CA · 92618-4210

Is this the product you have?

  • Clinical Chemistry, ABX Pentra Lipase CP, ABX Pentra 400/Pentrac400
  • REF
    • A11A01631
    • Model/Part # 1220001631
    • UDI # 03610230004944
Lot codes and dates
  • All Lots; All Software versions
Amount recalled
1539 Chemistry Analyzers

Where it was sold

Why it was recalled

Potential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could be observed and a patient result could be declared abnormal by the instrument due to these interferences instead of normal result < 38 U/L (or < reference value used in your laboratory).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Horiba Instruments Incorporated recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 16, 2020 · Data as of Sep 19, 4:49 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.