Reported Oct 30, 2024 · Ongoing

Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the…

Shimadzu Corporation Medical Systems Division

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION · Nishinokyo Kuwabara -Cho · Kyoto, N/A · N/A

Is this the product you have?

  • Trinias unity, Digital Angiography System with Catheterization Table
  • ModelKS-100 This system consists
  • of the following main componentsC arm, catherization table, image processing equipment, x-ray high voltage generator, x-ray tube. There are several types of each component
Lot codes and dates
  • UDI-DI
    • 04540217058327
    • 04540217049066
    • 04540217058143
  • Serial Numbers
    • 41EFCD7AC001
    • 41E58C773001
    • 41EFCD79C001
    • 41EFCD78B001
    • 41EFC73A9001
    • 41EFCD7D4001
    • 41E58C76B001
    • 41EFC73D9001
    • 41EFCD77C001
    • 41EFC73A5001
    • 41EFC73C5001
    • 41EFCD79A001
    • 41EFCD7A6001
Amount recalled
13

Where it was sold

Why it was recalled

If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Shimadzu Corporation Medical Systems Division recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:55 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.