Reported Oct 21, 2026 · Ongoing · Updated Oct 26, 2026

REF: P65 ST862 CoLag SCREW, FT, 6.7X62.5 mm

In2bones USA, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: In2bones USA, LLC · 7750a Trinity Rd · Cordova, TN · 38018-2722

Is this the product you have?

  • REF
    • P65 ST862 CoLag SCREW
    • FT
    • 6.7X62.5 mm
Lot codes and dates
  • UDI
    • 00817906023992/Lot: GM008380
    • GM007868
Amount recalled
29 units

Where it was sold

Why it was recalled

Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new In2bones USA, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Nov 4, 12:11 AM UTC