Reported Nov 4, 2015 · Terminated

Nipro Safe Touch TULIP Safety Fistula Needle, FT+152530TP, for use as a blood access device for blood purification and…

Nipro Medical Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Nipro Medical Corporation · 3150 Nw 107th Ave · Miami, FL · 33172

Is this the product you have?

  • Nipro Safe Touch TULIP Safety Fistula Needle, FT+152530TP, for use as a blood access device for blood purification and for other treatments requiring an extracorporeal circuit of larger volumes of blood.
Lot codes and dates
  • Lot # 15C03
Amount recalled
50,000 pieces

Where it was sold

Why it was recalled

Some 1 1/4" needles were packaged and labeled in the 1" blister package.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Nipro Medical Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 13, 2017 · Data as of Sep 19, 5:20 PM UTC