Reported Dec 6, 2017 · Terminated
BRAINLAB EXACTRAC VERO, Model/Catalog Numbers: 46228 EXACTRAC VERO 3.5 46238 EXACTRAC VERO 3.5 46216 EXACTRAC VERO SW…
Brainlab AG
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Brainlab AG · Olof-Palme-Strasse 9 · Munich, N/A · N/A
Is this the product you have?
- BRAINLAB EXACTRAC VERO
- Model/Catalog Numbers46228 EXACTRAC VERO 3.5 46238 EXACTRAC VERO 3.5 46216 EXACTRAC VERO SW UPDATE 3.5.2 TO 3.5.3 46218 EXACTRAC VERO SW UPDATE 3.5.3 TO 3.5.4 The ExacTrac Vero system is intended to be used in conjunction with the MHI-TM2000 Radiation Therapy Linear Accelerator System manufactured by Mitsubishi Heavy Industries, Ltd. ExacTrac Vero uses the images received from the MHI-TM2000 linear accelerator for analyzing the current patient position and calculating - when applicable - a necessary correction shift. The correction shift is then exported to the MHI-TM2000 linear accelerator. The ExacTrac Vero system uses stereoscopic X-ray or Cone Beam CT registration and optical tracking of infrared reflective markers in order to localize and correct the patient position before and during treatment. Optionally ExacTrac Vero provides position data for the pan/tilt motion of the MHI-TM2000 gantry head to the MHITM2000 controller for continuous alignment of the beam orientation with a moving target. The position data is based on target detection via X-ray imaging and IR tracking of external surrogate markers
- Lot codes and dates
- Potentially affected are ExacTrac Vero versions 3.x. Currently only ExacTrac Vero SW versions 3.5(.x) are installed for use on existing ExacTrac Vero Systems (only ExacTrac Vero v.3.5 exist in the field)
- UDI04056481132446
- Amount recalled
- 28 units
Why it was recalled
Brainlab has internally detected that under specific conditions ExacTrac Vero may not correctly account for this ring angle correction during the calculation of the corresponding couch shift. This may result in the treatment couch not moving to the exact planned position resulting in a deviation between the planned and the treatment target position. This deviation may not be clearly visible to the user if no verification image of the patient position is acquired after this couch shift has been performed.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Brainlab AG recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.