Reported Nov 3, 2021 · Terminated
Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid…
Ab Sciex
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: AB SCIEX · 71 Four Valley Dr · Concord, N/A · N/A
Is this the product you have?
- Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid chromatography-tandem mass spectrometry (LC-MS/MS) systems. Model No. 5088288
- Lot codes and dates
- Software version 3.4
- Amount recalled
- 30
Where it was sold
Why it was recalled
The values of the Internal Standard (IS) concentrations are incorrectly derived when the user builds a customized test, then set the IS column to hidden and left the column empty. This may lead to the user making incorrect conclusions using incorrect results.
This product could cause temporary or reversible health problems — stop using it.