Reported Oct 22, 2025 · Ongoing
Medical convenience kits Item Number/Description FHCP08Y C SECTION PK HICS50I C-SECTION PACK HSCB90B C-SECTION BASIN…
American Contract Systems Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: American Contract Systems Inc. · 1601 Se Gateway Dr Ste 120 · Grimes, IA · 50111-5223
Is this the product you have?
- Medical convenience kits Item Number/Description FHCP08Y C SECTION PK HICS50I C-SECTION PACK HSCB90B C-SECTION BASIN PACK
- Lot codes and dates
3 of 3 products
Products covered by this recall # Barcodes Product as published 1 191072232724FHCP08Y UDI-DI 191072232724 Lot 8216211 Bag serial number 68258296 2 191072212351HICS50I UDI-DI 191072212351 Lot 8161711 Bag serial number 68273637 3 191072141606HSCB90B UDI-DI 191072141606 Lot 8083511 Bag serial numbers 47506618 47508489 * If bag serial number is listed as "N/A", then whole lot is affected - Amount recalled
- 32,433 total
This recall covers 18 products
The agency filed one notice per product. You are reading one of them; the other 17 are below.
5 more are in the archive.
Where it was sold
Why it was recalled
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new American Contract Systems Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.