Reported Nov 19, 2014 · Terminated
Philips Medical Systems Allura Xper FD20C with software version R8.2.O System Code: 722035. The Allura Xper FD series…
Philips Medical Systems, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Medical Systems, Inc. · 3000 Minuteman Rd · Andover, MA · 01810-1032
Is this the product you have?
- Philips Medical Systems Allura Xper FD20C with software version R8.2
- O System Code722035. The Allura Xper FD series are intended
- for use on human patients to performVascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures
- Lot codes and dates
- Allura Xper Software version R8.2.0
- Amount recalled
- 2 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Faulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Medical Systems, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.