Reported Nov 20, 2013 · Terminated

Applied Medical's Inzii¿ 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is…

Applied Medical Resources Corp.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Applied Medical Resources Corp · 22872 Avenida Empresa # 3 · Rancho Santa Margarita, CA · 92688-2650

Is this the product you have?

  • Applied Medical's Inzii¿ 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is classified as a Class II device. It was cleared to market via 510(k) filing K060051. The retrieval bag is a disposable specimen retrieval bag intended for use as receptacle for the collection of tissue, organs, and calculi during laparoscopic procedures. Applieds CD004 has a shelf life of three years. It is provided sterile and packaged in a heat sealed Tyvek pouch.
Lot codes and dates
  • Model Number CD004
  • Lot Nubmers

    75 of 75 codes

    • 1161268
    • 1161269
    • 1164934
    • 1168358
    • 1168361
    • 1169317
    • 1169318
    • 1170694
    • 1170700
    • 1170701
    • 1171214
    • 1172482
    • 1172664
    • 1174287
    • 1174793
    • 1174857
    • 1175123
    • 1175280
    • 1175472
    • 1176555
    • 1177023
    • 1177871
    • 1179069
    • 1179392
    • 1179460
    • 1179775
    • 1179890
    • 1180291
    • 1180411
    • 1180576
    • 1181647
    • 1181648
    • 1181649
    • 1181939
    • 1182760
    • 1182934
    • 1183225
    • 1183823
    • 1183957
    • 1184273
    • 1184728
    • 1184915
    • 1184916
    • 1187779
    • 1188817
    • 1189117
    • 1189957
    • 1190294
    • 1190358
    • 1191208
    • 1192057
    • 1192969
    • 1193196
    • 1193661
    • 1194010
    • 1194666
    • 1194807
    • 1195518
    • 1195727
    • 1195997
Amount recalled
29,506 units

Where it was sold

  • Nationwide

Why it was recalled

Applied Medical is conducting a voluntary recall of the Inzii¿ 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the sterile barrier. The likelihood of this situation to occur and result in patient harm is highly unlikely; however, out of an abundance of caution for patient safety and a commitmen

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new packaged foods recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 15, 2014 · Data as of Sep 19, 6:03 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.