Reported Oct 24, 2018 · Terminated

Reinforced Dual Lumen Cannula, 32Fr., REF: VV32F, Sterile EO Indications: The OriGen Reinforced Dual Lumen Cannula is…

OriGen Biomedical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: OriGen Biomedical, Inc. · 7000 Burleson Rd Bldg D · Austin, TX · 78744-3202

Is this the product you have?

  • Reinforced Dual Lumen Cannula, 32Fr
  • REFVV32F
  • Sterile EO IndicationsThe OriGen Reinforced Dual Lumen Cannula is indicated for use as a single cannula for both venous drainage and re-infusion of blood in the internal jugular vein during extracorporeal life support procedures
Lot codes and dates
  • UDI00816203022148
  • LotP18806
Amount recalled
73

This recall covers 7 products

The agency filed one notice per product. You are reading one of them; the other 6 are below.

Where it was sold

Why it was recalled

These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new packaged foods recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 15, 2021 · Data as of Sep 19, 6:51 PM UTC