Reported Oct 28, 2020 · Terminated
7" (18 cm) Appx 0.31 ml, Smallbore Pressure Infusion (400psig) Ext Set w/Remv MicroClave Clear, Clamp, NanoClave…
ICU Medical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: ICU Medical, Inc. · 951 Calle Amanecer · San Clemente, CA · 92673-6212
Is this the product you have?
3 of 3 products
| # | Barcodes | Product as published |
|---|---|---|
| 1 | 7" (18 cm) Appx 0.31 ml, Smallbore Pressure Infusion (400psig) Ext Set w/Remv MicroClave Clear, Clamp, NanoClave T-Connector (Purple Ring), Rotating Luer | |
| 2 | 10887709088190 | unit per pouch, 50 multivac pouches per case. UDI:(01)10887709088190(17)241201(30)50(10)4515232 |
| 3 | 10887709068350 | (01)10887709068350(17)250101(30)50(10)4572761 The high pressure extension set is a device used as conduit tubing to deliver fluids from an ~f electronic or manual injector and may also be used as connector tubing to provide a sterile fluid pathway between two devices. |
- Lot codes and dates
- Lot Numbers
45152324572761
- Lot Numbers
- Amount recalled
- N/A
This recall covers 99 products
The agency filed one notice per product. You are reading one of them; the other 98 are below.
86 more are in the archive.
Where it was sold
Why it was recalled
Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new ICU Medical, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.