Reported Oct 31, 2018 · Terminated
Smith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral…
Smith & Nephew, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Smith & Nephew, Inc. · 1450 E Brooks Rd · Memphis, TN · 38116-1804
Is this the product you have?
- Smith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411
- QTY(1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to make bone cuts on the distal femur to prepare the bone to receive a LEGION femoral implant
- Lot codes and dates
- Lot 17LM05396
- Amount recalled
- 6 units
Where it was sold
Why it was recalled
One lot of the Legion AP femoral cutting block, size 6, was manufactured with an oversized spring.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Smith & Nephew, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.