Reported Dec 13, 2017 · Terminated
Dial-A-Dose Insulin Delivery Device (Pen-Injector) cartridge holders Product Usage: The NovoPen Echo¿ is a re-usable…
Novo Nordisk Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Novo Nordisk Inc · 800 Scudders Mill Rd · Plainsboro, NJ · 08536-1606
Is this the product you have?
- Dial-A-Dose Insulin Delivery Device (Pen-Injector)
- cartridge holders Product UsageThe NovoPen Echo¿ is a re-usable pen injector for single patient use by diabetics for the self injection of a desired dose of insulin. The pen injector uses Penfill 3 mL cartridge of NovoLog¿, 100 units/mL (Ux100) [insulin aspart (rDNA origin) injection] and a single use detachable and disposable pen needle (supplied separately). The pen injector allows the user to dial the desired dose from 0.5 to 30 units in 0.5 increments
- Lot codes and dates
- U.S
- batch numbers
EVG1221EVG1226FVG7149FVG7458FVG8134 & FVG8135
- Amount recalled
- 16,557 units
Where it was sold
- Nationwide
Why it was recalled
A small number of batches may crack or break if exposed to certain chemicals, like certain cleaning agents.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Novo Nordisk Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.