Reported Oct 29, 2025 · Ongoing

Salem Sump PVC Tubes: Product Code/Product Name: 264929 Salem Sump" Stomach Tube, Dual Lumen, 12 Fr/Ch (4.0 mm) 264945…

Cardinal Health 200, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Cardinal Health 200, LLC · 3651 Birchwood Dr · Waukegan, IL · 60085-8337

Is this the product you have?

  • Salem Sump PVC TubesProduct
  • Code/Product Name264929 Salem Sump" Stomach Tube, Dual Lumen, 12 Fr/Ch (4.0 mm) 264945 Salem Sump" Stomach Tube, Dual Lumen, 14 Fr/Ch (4.7 mm) 264960 Salem Sump" Stomach Tube, Dual Lumen, 16 Fr/Ch (5.3 mm) 264986 Salem Sump" Stomach Tube, Dual Lumen, 18 Fr/Ch (6.0 mm) 5558264957 Salem Sump" COILED TUBE 16Fr/Ch (5.3 mm) 5558264965 Salem Sump" TUBE 16 Fr/Ch (5.3 mm) 5558264973 Salem Sump" COILED TUBE 18Fr/Ch (6.0 mm) 5558264981 Salem Sump" TUBE 18 Fr/Ch 18Fr/Ch (6.0 mm) 8888264911 Salem Sump" Dual Lumen Stomach Tube, 10 Fr/Ch (3.3 mm), 36" (91 cm) 8888268060 Salem Sump" Dual Lumen Stomach Tube 6 Fr/Ch (2.0 mm), 24" (61 cm) 8888268086 Salem Sump" Dual Lumen Stomach Tube, 8 Fr/Ch (2.7 mm), 24" (61 cm)
Lot codes and dates
  • Product Code/Product Name/UDI-DI264929 Salem Sump" Stomach Tube, Dual Lumen, 12 Fr/Ch (4.0 mm) 10884521067196 264945 Salem Sump" Stomach Tube, Dual Lumen, 14 Fr/Ch (4.7 mm) 10884521067202 264960 Salem Sump" Stomach Tube, Dual Lumen, 16 Fr/Ch (5.3 mm) 10884521067219 264986 Salem Sump" Stomach Tube, Dual Lumen, 18 Fr/Ch (6.0 mm) 10884521067226 5558264957 Salem Sump" COILED TUBE 16Fr/Ch (5.3 mm) 50192253008171 5558264965 Salem Sump" TUBE 16 Fr/Ch (5.3 mm) 50192253008195 5558264973 Salem Sump" COILED TUBE 18Fr/Ch (6.0 mm) 50192253008201 5558264981 Salem Sump" TUBE 18 Fr/Ch 18Fr/Ch (6.0 mm) 50192253008218 8888264911 Salem Sump" Dual Lumen Stomach Tube, 10 Fr/Ch (3.3 mm), 36" (91 cm) 10192253012453 8888268060 Salem Sump" Dual Lumen Stomach Tube 6 Fr/Ch (2.0 mm), 24" (61 cm) 10192253012972 8888268086 Salem Sump" Dual Lumen Stomach Tube, 8 Fr/Ch (2.7 mm), 24" (61 cm) 10192253012996 All Lots
Amount recalled
all lots

This recall covers 6 products

The agency filed one notice per product. You are reading one of them; the other 5 are below.

Where it was sold

Why it was recalled

Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Cardinal Health 200, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:15 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.