Reported Oct 31, 2018 · Terminated
Mitek Screw and Washer Depth Gauge, Product Code 219024
DePuy Mitek, Inc., a Johnson & Johnson Co.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: DePuy Mitek, Inc., a Johnson & Johnson Co. · 325 Paramount Dr · Raynham, MA · 02767-5199
Is this the product you have?
- Mitek Screw and Washer Depth Gauge, Product Code 219024
- Lot codes and dates
- UDI10886705005390
- Lot Numbers
00051781103107000910111051320009181111210601021571201104010304595022902053559612320207451970198020859797122903016159805850302184980912103033379907650311333C0000076550311389C0000125270512552C000020182060557110001968420612501100041574907075121000431580071050110006617130801536100111999608091171002144090513010061130912100100812010000645141103103
- Amount recalled
- 401
Where it was sold
Why it was recalled
The device has an offset in the design that results in a reading approximately 3.2mm shorter than the actual length of the bone hole.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new DePuy Mitek, Inc., a Johnson & Johnson Co. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.