Reported Nov 8, 2023 · Ongoing
Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator
Abbott
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Abbott · 5050 Nathan Ln N · Plymouth, MN · 55442-3209
Is this the product you have?
- Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator
- Lot codes and dates
- UDI/DI 05415067022417
- serial numbers
16930509169833301758340717634837177839281778393017905961179284131795452117991999180473841849371118522937186750681869092018709734187429972475-122504-122547-132885-143003-143278-153450-153550-15179423571800331018102335181320171813209218657521186666471918211919197174
- Amount recalled
- 34 devices
Why it was recalled
Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.
This product could cause temporary or reversible health problems — stop using it.