Reported Dec 13, 2017 · Terminated
SMARTABLATE RF System Generator
Biosense Webster, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Biosense Webster, Inc. · 15715 Arrow Hwy · Irwindale, CA · 91706-2006
Is this the product you have?
- SMARTABLATE RF System Generator
- Lot codes and dates
Product Codes M490006M490007
- Amount recalled
- 185 units total
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
Why it was recalled
Generator may inadvertently deliver RF energy if the foot pedal is placed in close proximity and oriented toward the Stereotaxis magnets.
This product could cause temporary or reversible health problems — stop using it.