Reported Nov 19, 2014 · Terminated

Masimo SET uSpO2 Cable The Masimo SET uSpO2 Pulse Oximetry Cable is an in-line cable that provides continuous…

Masimo Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Masimo Corporation · Forty Parker · Irvine, CA · 92618

Is this the product you have?

  • Masimo SET uSpO2 Cable The Masimo SET uSpO2 Pulse Oximetry Cable is an in-line cable that provides continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and pulse rate when installed in a compatible OEM host system
  • Shipping/PackageAll Pulse Oximetry cables are shipped non-sterile in a cardboard carton. They are intended to be used non-sterile
Lot codes and dates
  • PART_ID LOT_ID

    140 of 140 codes

    • 3311
    • 11CE2
    • 3311
    • E11EA7
    • 3311
    • 11EMM
    • 3311
    • E11EDJ
    • 3311
    • 11FAU
    • 3311
    • 11FAV
    • 3311
    • 11CE2
    • 3311
    • 11FAW
    • 3311
    • 11GBY
    • 3311
    • 11HGF
    • 3311
    • 11HE6
    • 3311
    • 11J80
    • 3311
    • 11JCJ
    • 3311
    • 11HNO
    • 3311
    • 11JCK
    • 3311
    • 11K37
    • 3412
    • 11MA7
    • 3311
    • R11KFY
    • 3311
    • 11K37
    • 3311
    • 11KQM
    • 3311
    • 11NDZ
    • 3311
    • 12B95
    • 3412
    • 12CP1
    • 3412
    • 11AM7
    • 3449
    • E12HR3
    • 3450
    • E12HR4
    • 3412
    • 12CP1
    • 3635
    • 12KP4
    • 3636
    • 12KP5
    • 3635
    • 12KQ7
    PART_ID LOT_ID SERIAL_NUMBER

    1,878 of 1,878 codes

    • 3449
    • 2062190-001
    • 3315
    • E12JS1
    • 3449
    • V25060
    • 3449
    • V25061
    • 3449
    • V25062
    • 3449
    • V25063
    • 3449
    • V25064
    • 3449
    • V25065
    • 3449
    • V25066
    • 3449
    • V25067
    • 3449
    • V25068
    • 3449
    • V25069
    • 3449
    • V25070
    • 3449
    • V25071
    • 3449
    • V25072
    • 3449
    • V25073
    • 3449
    • V25074
    • 3449
    • V25075
    • 3449
    • V25076
    • 3449
    • V25077
    • 3449
    • V25078
    • 3449
    • V25079
    • 3449
    • V25080
    • 3449
    • V25081
    • 3449
    • V25082
    • 3449
    • V25083
    • 3449
    • V25084
    • 3449
    • V25085
    • 3449
    • V25086
    • 3449
    • V25087
Amount recalled
21,754 units (21,222) in the U.S. and Puerto Rico)

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Firm has identified a small number of Oximetry Cables with crossed-internal wires. The effect on performance when these wires are crossed is that when a low SpO2 value is measured, a high SpO2 value would be displayed on the monitoring device. Additionally when a high SpO2 value is measured, a low SpO2 value would be displayed on the monitoring device.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Masimo Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 28, 2016 · Data as of Sep 19, 6:54 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.