Reported Dec 20, 2017 · Terminated

Glidesheath Slender Hydrophilic Coated Introducer Sheath A-Kit, 5FR - 10CM (0.025), REF/Product Code RM*ES5J10HAT,…

Terumo Medical Corp.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Terumo Medical Corp · 2101 Cottontail Ln · Somerset, NJ · 08873-1277

Is this the product you have?

  • Glidesheath Slender Hydrophilic Coated Introducer Sheath A-Kit, 5FR - 10CM (0.025), REF/Product Code RM*ES5J10HAT, STERILE, Rx ONLY
Lot codes and dates
    • Lot 161110
    • 161214
    • 170119
Amount recalled
N/A

This recall covers 17 products

The agency filed one notice per product. You are reading one of them; the other 16 are below.

4 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

An incomplete seal at the chevron end of the primary packaging of the GLIDESHEATH SLENDER Introducer Sheath was detected in a small quantity of the affected lot population, 0.01% (1 in 10, 000 units). This breach of the sterile barrier may compromise the sterility of the sheath.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new packaged foods recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 25, 2019 · Data as of Sep 19, 6:54 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.