Reported Nov 6, 2019 · Terminated

SORIN Heater-Cooler System 3T 120V~/60Hz REF 16-02-85 Description of product: The St¿ckert Heater-Cooler System is an…

LivaNova USA Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: LivaNova USA Inc. · 14401 W 65th Way · Arvada, CO · 80004-3503

Is this the product you have?

  • SORIN Heater-Cooler System 3T 120V~/60Hz
  • REF 16-02-85 Description of productThe St¿ckert Heater-Cooler System is an independent (i.e. independent of the water supply) 3-circuit-heating/cooling system. The three separate water circuits
  • have been developed to serve as" Two circuits for the patient supplied by one tank (for the heating/cooling blanket and the heat-exchanger associated with the oxygenator), common temperatures ranging from 2 degrees Celsius to 41 degrees Celsius, tank volume 6 liters. " One interchangeable heating/cooling circuit for cardioplegia heat-exchanger, temperatures ranging from 2 degrees Celsius to 10 degrees Celsius (cooling tank) and/or from 15 degrees Celsius to 41 degrees Celsius (heating tank). If required, the circuits for the patient and the circuit for cardioplegia can be switched off separately, in order to increase the activated functional group s heating and/or cooling performance
  • Product UsageThe Stockert Heater-Cooler System 3T is intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours
Lot codes and dates
  • 16-02-80 Stockert Heater-Cooler System 3T, 230V, Serial #s 16S10027 thru 16S16874 16-02-81 Stockert Heater-Cooler System 3T, 240 v/60 Hz, Serial #s 16S10743 thru 16S11708 16-02-82 Stockert Heater-Cooler System 3T, 208 v/60 Hz, Serial #s 16S10772 thru 16S16840 16-02-83 Stockert Heater-Cooler System 3T, 127V/60Hz, Serial #s 16S11455 thru 16S16509 16-02-85 Stockert Heater-Cooler System 3T, 120 v/60 Hz, Serial #s 16S10958 thru 16S16847 16-02-95 Stockert Heater-Cooler System 3T, 200V/50Hz/60Hz, Serial #s 16S12004 thru 16S16818
Amount recalled
8671

Where it was sold

Why it was recalled

The firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. In addition, the firm is releasing a design upgrade to reduce the risk of potential emission of aerosols from the 3T.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new LivaNova USA Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 4, 2022 · Data as of Sep 19, 6:55 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.