Reported Nov 6, 2024 · Ongoing
CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD…
Smiths Medical ASD Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Smiths Medical ASD Inc. · 6000 Nathan Ln N · Minneapolis, MN · 55442-1690
Is this the product you have?
- CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following
- CADD Solis Ambulatory Infusion Systems
183 of 183 codes
21-2101-0100-5021-2101-0200-0221-2101-0200-0321-2101-0200-0621-2101-0200-0721-2101-0200-1221-2101-0200-1421-2101-0200-1721-2101-0200-23821-2101-24921-2101-5121-2102-0200-0221-2102-24921-2102-5121-2111-0100-0021-2111-0100-5021-2111-0100-5121-2111-0200-0221-2111-0200-0321-2111-0200-0621-2111-0200-0721-2111-0200-0821-2111-0200-1221-2111-0200-1421-2111-0200-1721-2111-0200-23221-2111-0200-5021-2111-0300-0021-2111-0300-0121-2111-0300-0221-2111-0300-0321-2111-0300-0621-2111-0300-0721-2111-0300-0821-2111-0300-0921-2111-0300-1221-2111-0300-1421-2111-0300-1721-2111-0300-23121-2111-0300-23221-2111-0300-5021-2111-0400-0121-2111-0400-5121-2111-0401-02L21-2111-0401-03L21-2111-0401-06L21-2111-0401-0721-2111-0401-07L21-2111-0401-08L21-2111-0401-12L21-2111-0401-14L21-2111-0401-17L21-2111-0401-24921-2111-0401-5121-2111-0401-51L21-2111-0401-7821-2111-0402-0221-2111-0402-02L21-2111-0402-03L21-2111-0402-06L
- Lot codes and dates
- All serial numbers.
- Amount recalled
- 138039 units
Why it was recalled
Stop using this product now
Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Smiths Medical ASD Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.