Reported Nov 16, 2022 · Ongoing

ORISE Gel Submucosal Lifting Agent--Indicated for use in gastrointestinal endoscopic procedures for submucosal lift of…

Boston Scientific Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Boston Scientific Corporation · 100 Boston Scientific Way · Marlborough, MA · 01752-1234

Is this the product you have?

  • ORISE Gel Submucosal Lifting Agent--Indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early stage cancers or other gastrointestinal mucosal lesions, prior to excision with a snare or other suitable
  • endoscopic device Packaged as followsUPN: (1)Twin Pack Box 1 M00519200 (2)Twin Pack Box 10 M00519201 (3)Single Pack Box 1 M00519220 (4)Single Pack Box 10 M00519221
Lot codes and dates
  • GTIN Number(1) 08714729974567 (2) 08714729974574 (3) 08714729993834 (4) 08714729993841 All batches
Amount recalled
402,968 Total

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Boston Scientific Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice