Reported Oct 28, 2026 · Ongoing · Updated Nov 2, 2026

Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/105; Material Number: M001195290

Boston Scientific Cork, Ltd.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Boston Scientific Cork, Ltd. · Boston Scientific Limited, Business And Technology Park, Model Farm Road · Cork · N/A

Is this the product you have?

  • Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/105
  • Material NumberM001195290
Lot codes and dates
  • Material NumberM001195290
  • UDI-DI08714729839408
  • Batch Numbers
    • 38362172
    • 38362173
    • 38530450
    • 38595659
Amount recalled
42

This recall covers 5 products

The agency filed one notice per product. You are reading one of them; the other 4 are below.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Boston Scientific Cork, Ltd. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Nov 4, 12:15 AM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.