Reported Nov 23, 2022 · Ongoing
Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN
Getinge USA Sales Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Getinge Usa Sales Inc · 1 Geoffrey Way · Wayne, NJ · 07470-2035
Is this the product you have?
- Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN
- Lot codes and dates
- No UDI - product discontinued in 2008. Part Number 1140.20F0: Serial Nos. 3, 8, 9, 10 Part Number 1140,20AN: Serial No. 851
- Amount recalled
- 5 US; 1008 OUS
Where it was sold
Why it was recalled
The back rest may unintentionally drop suddenly, resulting in adverse health consequences if a patient were to fall on the floor head first and at a steep angle.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Getinge USA Sales Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.