Reported Nov 23, 2022 · Ongoing

Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN

Getinge USA Sales Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Getinge Usa Sales Inc · 1 Geoffrey Way · Wayne, NJ · 07470-2035

Is this the product you have?

  • Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN
Lot codes and dates
  • No UDI - product discontinued in 2008. Part Number 1140.20F0: Serial Nos. 3, 8, 9, 10 Part Number 1140,20AN: Serial No. 851
Amount recalled
5 US; 1008 OUS

Where it was sold

Why it was recalled

The back rest may unintentionally drop suddenly, resulting in adverse health consequences if a patient were to fall on the floor head first and at a steep angle.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Getinge USA Sales Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:50 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.