Reported Dec 20, 2017 · Terminated
XCELA PASV 5F DL 55CM IR-145 Nitinol Wire KIT PG, UPN H965251260, Catalog No. 25-126 The Xcela with PASV, BioFlo PICC…
Angiodynamics Inc. (Navilyst Medical Inc.)
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Angiodynamics Inc. (Navilyst Medical Inc.) · 10 Glens Falls Tech Park · Glens Falls, NY · 12801-3864
Is this the product you have?
- XCELA PASV 5F DL 55CM IR-145 Nitinol Wire KIT PG, UPN H965251260, Catalog No. 25-126 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.
- Lot codes and dates
- Lot 4904007
- Amount recalled
- 4591 units in total
This recall covers 13 products
The agency filed one notice per product. You are reading one of them; the other 12 are below.
Where it was sold
- Nationwide
Why it was recalled
Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Angiodynamics Inc. (Navilyst Medical Inc.) recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.