Reported Nov 14, 2012 · Terminated
STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and…
Diagnostica Stago, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Diagnostica Stago, Inc. · 5 Century Dr · Parsippany, NJ · 07054-4607
Is this the product you have?
- STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
- Lot codes and dates
- K082248 D089976 Ref 58104 units
- Amount recalled
- 232 units US; 5 units Canada
Where it was sold
- Nationwide
Why it was recalled
Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite¿.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Diagnostica Stago, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.