Reported Nov 14, 2012 · Terminated

Sysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan;…

Sysmex America, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Sysmex America, Inc. · 1 Nelson C White Pkwy · Mundelein, IL · 60060-9528

Is this the product you have?

5 of 5 products

Products covered by this recall
Model / numberProduct as published
Sysmex XE-5000 Automated Hematology System
an automated differential cell counter
Sysmex Corporation, Kobe, Japan
Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060
063-7591-7The Sysmex XE-5000 is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XE-5000 performs analyses using the following methods: RF/DC Detection Method, Sheath Row DC Detection Method, and Flow Cytometry Methods using a Semiconductor Laser. Particle characterization and identification is based on detection of forward scatter, fluorescence and adaptive cluster analysis. Using the same reagents as the XE-2100, the XE-5000 automatically classifies cells from blood and body fluids and carries out all processes automatically from aspiration of the sample to outputting the results.
Lot codes and dates
  • catalog #063-7591-7, all serial numbers
Amount recalled
763 units

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Sysmex America, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 22, 2016 · Data as of Sep 19, 6:15 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.