Reported Nov 14, 2012 · Terminated
R-30H , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-ray tubes 0.6/1.2P364DK-85,…
Shimadzu Medical Systems
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Shimadzu Medical Systems · 20101 S Vermont Ave · Torrance, CA · 90502-1328
Is this the product you have?
- R-30H , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-ray tubes 0.6/1.2P364DK-85, 0.6/1.2P324DK-125, 0.6/1.2P324DK-85, 0.6/1.2P38DE-85 Device Model # R-30H
- Subsequent Product CodeIZX
- Product UsageThe device is the beam-limiting device for X-ray Radiography
- Lot codes and dates
- Device Model #R-30H
- Amount recalled
- 665 units
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
- Nationwide
Why it was recalled
Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Shimadzu Medical Systems recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.