Reported Nov 14, 2012 · Terminated

Hoffman LRF Telescopic Strut Short Stryker Stryker Trauma AG CH-2545 Seizach Distributed in the USA by Howmedica…

Stryker Howmedica Osteonics Corp.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Stryker Howmedica Osteonics Corp. · 325 Corporate Drive · Mahwah, NJ · 07430

Is this the product you have?

  • Hoffman LRF Telescopic Strut Short Stryker Stryker Trauma AG CH-2545 Seizach Distributed in the USA by Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430. The use of external devices provides the surgeon with a means of bone fixation for the management of fracture and reconstructive surgery. These devices are intended only to assist healing and are not intended to replace normal bone structures. External fixation devices are intended to hold fractured bone surfaces in apposition to facilitate normal healing. While proven successful, the devices are manufactured from metal, plastic, and composite materials. No such fracture fixation device, subject to fatigue, can be expected to withstand activity levels in the same way as would a normal health bone. The external fixation system, therefore, will not be as strong, reliable or durable as a normal human bone.
Lot codes and dates

22 of 22 products

Products covered by this recall
#Product as published
1K113327 Catalog No: Telescopic Strut Extra Short Hoffman LRF Length: 4933-0-100 :100-125mm (Black) 4933-0-120 Telescopic Strut Short Hoffman LRF Length: 119-161 mm (Yellow) 4933-0-140 Telescopic Strut Medium Hoffman LRF Length 138-201 mm (Blue) 4933-0-180 Telescopic Strut Long Hoffman LRF Length: 177-277 mm (Red)
2Lot Code: 4933-0-100 Telescopic Strut Extra Short Hoffman LRF Length 100-125mm (Black) X28937
3Z04197
4Z09163
5Z09597
6Z12135
74933-0-120 Telescopic Strut Short Hoffman LRF Length: 119-161 mm (Yellow) X28961
8Z04199
9Z09161
10Z09769
11Z12012
124933-0-140 Telescopic Strut Medium Hoffman LRF Length 138-201 mm (Blue) X34352
13Z04200
14Z05068
15Z09011
16Z09598
17Z10872
184933-0-180 Telescopic Strut Long Hoffman LRF Length: 177-277 mm (Red)
19X28963
20Z04429
Amount recalled
165 units

Where it was sold

  • Nationwide

Why it was recalled

Three complaints have been filed where the Hoffman LRF Telescopic Struts have broken during load bearing application by patients having a body weight greater than 250 LBS.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Stryker Howmedica Osteonics Corp. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 17, 2014 · Data as of Sep 19, 4:51 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.