Reported Nov 25, 2015 · Terminated
Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15…
Precision Medical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Precision Medical, Inc. · 300 Held Dr · Northampton, PA · 18067-1150
Is this the product you have?
- Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.
- Lot codes and dates
- Lot numbers 072215 and 072315.
- Amount recalled
- 315
Where it was sold
Why it was recalled
These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).
This product could cause temporary or reversible health problems — stop using it.